University of Saskatchewan • College of Agriculture and Bioresources • Department of Soil Science Lab 5E19 • lab5e19@gmail.com
Department of Soil Science • USask

Greenhouse Gas, Biogas & Isotope Analysis Services

The Prairie Environmental Agronomy Research Laboratory (PEARL)—housing the agricultural Greenhouse Gas Analysis Laboratory (agGAL)—provides high-throughput gas chromatography and cavity ring-down spectroscopy for researchers at the University of Saskatchewan and partner institutions across Canada.

4 Gases Included N2O, CO2, CH4 & O2 in standard GHG analysis
Supplies Provided Evacuated Exetainers, labels, syringes & needles on request
2–4 Weeks Recommended window to fill freshly evacuated vials
100–400 Vials Recommended batch size (surcharge applies under 100)
Analytical Capabilities

Gas Analysis Services & Pricing

Rates shown reflect standard Canadian academic pricing. Contact PEARL in advance for high-concentration samples, large projects, or custom analytical sequences. Full department fee schedules are also published on the USask Soil Science Facilities page.

Most Popular • Soil & Chamber Fluxes

Standard Greenhouse Gas Analysis (GC)

$4.00 CAD / sample

High-throughput simultaneous trace-gas quantification for static chamber studies, soil incubations, CO2 burst assays, and soil profile sampling.

N2O CO2 CH4 O2
  • Instrument: Bruker 450-GC (ECD, FID, TCD)
  • Detection limits: ~0.03 ppm N2O, ~0.05 ppm CH4, ~25 ppm CO2
  • Injection volume: 20 mL into evacuated 12 mL Exetainer (min. 18 mL)
  • Extended range: Calibratable up to 100% N2O for soil profile samples (notify in advance)
Livestock & High Concentration

Biogas & Ruminal Fermentation Analysis

$8.00 CAD / sample

Dedicated high-concentration gas analysis for in vitro ruminal fluid incubations, RUSITEC trials, enteric methane studies, and biogas samples.

CH4 CO2 H2 O2 N2
  • Instrument: Agilent 490 Micro GC
  • Injection volume: ≥12 mL for 5.9–6.8 mL vials; ≥18 mL for 12 mL vials
  • Advance planning required: Uses a separate GC requiring setup & custom calibration standards
  • Detector protection: Always disclose expected high CH4 / CO2 concentrations beforehand
Stable Carbon Isotope • CRDS

δ13C-CO2 Isotope Analysis

$14.00 CAD / sample

Cavity Ring-Down Spectroscopy (CRDS) for quantifying δ13C-CO2 and CO2 concentration in soil respiration, substrate-labeling (e.g., 13C-glucose), and decomposition studies.

δ13C-CO2 CO2 (ppm)
  • Instrument: Picarro CRDS Analyzer
  • Minimum concentration: Sample CO2 must be > 300 ppm
  • Injection volume: Inject 30 mL for accurate CO2 concentration + δ13C-CO2 (12 mL suffices for δ13C-CO2 only)
  • Enriched samples: Flag 13C-enriched vs. natural-abundance samples on your sheet to optimize run order
Nitrogen Isotopomers & Bulk 15N

15N2O Isotopomers & Partner IRMS

Quote Confirm availability

Isotopic N2O pathway partitioning (δ15Nα, δ15Nβ, site preference, δ18O) via CRDS, plus coordination with the Department's Stable Isotope Lab for bulk 15N-N2O and 15N-N2.

δ15N2O δ15Nα / δ15Nβ δ18O 15N2 (IRMS)
  • CRDS Instrument: Picarro G5131-i (confirm operational status & precision before sampling)
  • Partner IRMS Facility: Soil Science Stable Isotope Lab (isoFlow GHG IRMS, Room 5E80)
  • Note on precision: Small septum-sealed vials with variable δ15N have higher real-world noise than continuous-tank specs
Service Analytes Rate (Academic) Recommended Vial & Injection Key Requirement
Standard GHG (GC) N2O, CO2, CH4, O2 $4.00 / sample 12 mL Exetainer; inject 18–20 mL (5.9 mL vial: inject 12 mL) Extra fee for batches < 100 samples or extreme concentrations
Biogas / Rumen (Micro GC) H2, O2, N2, CH4, CO2 $8.00 / sample 5.9–6.8 mL vial: ≥12 mL; 12 mL vial: ≥18 mL Schedule ahead; state expected CH4 & CO2 levels
δ13C-CO2 (CRDS) δ13C-CO2, CO2 $14.00 / sample 12 mL Exetainer; inject 30 mL (or 12 mL for δ13C only) CO2 must exceed 300 ppm; select "Custom: 13CO2" on form
iN2O / Bulk 15N 15N2O isotopomers; 15N2 Request quote 12 mL double-septa Exetainer Confirm Picarro G5131-i / IRMS status before starting trial

What Is Included — and What Is Additional?

  • Included in per-sample cost: Freshly evacuated 12 mL Labco Exetainer® vials (double septa), removable Avery/laser labels, and syringes & needles upon request.
  • Additional costs: Off-campus shipping (approx. $40 per box of 400 vials via Purolator Express), small-batch surcharge for submissions under 100 samples, surcharges for out-of-range concentrations, and replacement fees for unreturned or missing vials.
  • Flux calculations: PEARL reports gas concentrations. Flux and cumulative emission calculations are the researcher's responsibility (additional fee if requested). See Venterea et al. (2020) for standard chamber flux methodology.
Submission Workflow

How to Prepare & Submit Gas Samples

Following these five steps prevents autosampler jams, sample mix-ups, leaky vials, and billing delays. Please contact lab5e19@gmail.com before your experiment begins.

1

Plan Ahead & Request Evacuated Vials

Email lab5e19@gmail.com with your expected sample count, target gases, concentration range, and sampling dates.

  • Lead time: Give at least 1 day notice for USask campus pickup (placed outside Room 5E31) and 1 week notice for off-campus shipments (Purolator Express arrives in 1–3 business days).
  • Order 5% spare vials: Always request ~5% extra vials to cover breakage or sampling errors.
  • Stagger multi-month studies: Evacuated vials have a limited vacuum shelf life and should be filled within 2–4 weeks. For long incubations, request shipments in staggered batches every 2–4 weeks rather than all at once.
  • Courier details: External clients must provide a full institutional shipping address and a recipient phone number for the courier.
2

Collect Samples with Proper Technique

  • Needle gauge: Use 22G to 25G needles (PEARL recommends 22G for optimal flux linearity). Avoid ≥21G needles (they tear septa, causing selective-membrane leaks) and 26G needles (too restrictive).
  • Wait before withdrawing: After pulling back the syringe plunger, pause at least 5 seconds with 22G (or 10 seconds with 25G) so gas finishes equilibrating into the syringe.
  • Over-pressurize the vial: Inject 18–20 mL into 12 mL Exetainers for standard GHG analysis, 12 mL into 5.9–6.8 mL vials, or 30 mL for δ13C-CO2 + accurate CO2 concentrations. Extra pressure allows re-analysis if needed.
  • Include ambient air controls: Collect at least 8 atmospheric/ambient air samples per sampling event to verify background baselines and detect site variability.
3

Label Every Vial Correctly

  • Use ONLY PEARL-provided labels: Our labels fit the GC autosampler racks and wash off cleanly for vial reuse. Print using Microsoft Word template Avery 8167 (or Avery 15667), or write legibly by hand.
  • If supplying your own labels: Use a single layer of clear plastic laser labels—OnlineLabels OL25CX ("Clear Matte for Laser")—printed with a laser printer.
  • Placement & single layer only: Place each label close to the cap. Never stack labels on top of one another (stacked or thick paper labels jam automated racks). If relabeling is unavoidable, apply the new label beside the old one.
  • Unique, descriptive IDs: Use an unmistakable format such as 2026Jan14_Elora_plot101_T15 so any misplaced vial can still be identified.
4

Organize Boxes in Continuous Order & Complete the Request Form

  • Continuous order, NO gaps: Arrange vials sequentially left-to-right, row-by-row, across boxes 1 to N with no empty spaces.
  • Exact 1:1 match with Excel form: Email a completed and signed GHG Analysis Request Form (Excel) listing only the samples actually in the boxes, in the exact physical order they are packed. Remove any uncollected or skipped samples from the sheet—missing vials left on the form will be billed.
  • Label trays and box orientation: Place tape on each tray matching the Sample # range from the form. Mark the starting corner (e.g., "Top-left corner, left-to-right, row-by-row") and record the 4 corner vial IDs (top-left, top-right, bottom-left, bottom-right) to verify orientation.
  • Exterior box label: Clearly write your name, university/department, contact email/phone, date, box number (1 of N), and analysis type on the outside of every box (never just "Samples").
5

Ship or Drop Off & Confirm Billing

  • Storage before shipping: Filled Exetainers hold reliably for ~1 month; ship batches of 100–400 samples promptly to minimize leak outliers and avoid the <100-sample surcharge.
  • Email in advance: Notify lab5e19@gmail.com with your completed Request Form and tracking number before shipping (check holiday/seasonal closures first).
  • USask campus drop-off: Bring trays to Room 5E19 (or leave outside the door of Room 5E15 if no one is in 5E15, 5E19, or 5E31) and include your CFOAPAL account number on the form.
  • External institutions: Ship to Lab 5E19 (address below) and arrange a Purchase Order made out to "University of Saskatchewan". Return all PEARL-supplied vials (including unused spares) to avoid missing-vial charges.

Box Organization Visual Reference

Pack vials left-to-right, row-by-row with zero gaps, and label each tray with its matching form range and corner IDs.

Partial Shipments & Missing Vials

Do not leave empty gaps in trays to reserve spots for future sampling dates. Pack each shipment continuously and submit an Excel Request Form matching only the vials in that shipment.

Interactive Pre-Submission Checklist

Verify each item before dropping off or couriering your samples.

0 of 8 items completed
Lessons from Real Submissions

Technical Tips & Troubleshooting Guide

Compiled directly from years of sample submissions from USask, University of Manitoba, and University of Guelph research teams. Avoiding these common issues protects your data quality and turnaround time.

Vial Dimensions & Double Septa

  • Standard autosampler racks require 12 mL Labco Exetainers (15.6 mm OD × ~101 mm tall with cap) fitted with double septa for reliable sealing.
  • 16 mm OD vials do NOT fit our GC racks.
  • Labco 5.9–6.8 mL vials can be accommodated, but notify the lab in advance (especially if they carry large paper labels) so custom racks can be prepared.
  • Avoid Labco pre-evacuated vials: Factory-evacuated vials are not suitable for trace-gas research. Let PEARL evacuate vials in-house.

Cap Tightness & Diagnosing Leaks

  • Check caps BEFORE evacuation: A loose cap feels tight under vacuum, then loosens when pressurized with sample gas—causing leaks and lab-air readings.
  • Do not overtighten: Overtightening causes the septum to bulge outward and fail.
  • Diagnostic clue: Unusually high CH4 (~3–4 ppm) in soil/ambient samples often indicates a vial that was left under vacuum or had a loose cap.
  • Never wrap plumber's tape around vial caps—it does not stop leaks and interferes with autosampler trays.

Needle Gauge & Syringe Patience

  • Use 22G–25G needles; 22G is PEARL's preferred field gauge for cleaner 4-point flux linearity.
  • Avoid ≥21G needles: Oversized needles punch holes in septa that act as selective membranes, artificially elevating O2 and N2O.
  • Pause after pulling the plunger: Hold a syringe to your ear in a quiet room—air keeps flowing for ~5 s (22G) to ~10 s (25G). Wait before withdrawing!
  • If operating your own evacuation manifold, use 23.5G needles and replace them daily.

Removing & Replacing Labels

  • Never stick a new label over an old one—two layers make the vial too wide for GC autosampler trays.
  • If an unanalyzed vial needs a corrected label, apply the second label beside the original label, not on top.
  • To clean reusable vials, simply submerge them in a bucket of soapy water for ~12 hours. Because properly capped Exetainers are airtight, water is harmless and labels slide off cleanly.

High-N2O Carry-Over & Outliers

  • N2O is chemically "sticky" compared to CO2 and CH4. Simple evacuation does not eliminate carry-over from high-concentration vials.
  • PEARL removes caps from any vial exceeding 10 ppm N2O immediately after GC analysis and airs them out overnight before reuse.
  • If one timepoint in a 4-point chamber flux jumps abruptly (e.g., 0.34 → 0.34 → 0.34 → 1.38 ppm while CO2 rises linearly), suspect a contaminated/reused vial outlier.

QA/QC Controls vs. DIY Standards

  • Collect at least 8 ambient/atmospheric air samples per sampling event (especially near barns or livestock) to establish robust background baselines.
  • DIY cylinder standards are discouraged: Unverified lab cylinders labeled "1 ppm N2O" often actually measure 0.8–1.2 ppm and create false confusion, while vial leakage is random rather than systematic.
  • PEARL calibrates every run using certified multi-point standards (Praxair / Linde).

✔ Best Practices (Do)

  • Contact lab5e19@gmail.com before starting your trial and order 5% spare vials.
  • Fill evacuated vials within 2–4 weeks and ship batches of 100–400 vials within ~1 month.
  • Use 22G–25G needles and wait 5–10 seconds after pulling the syringe plunger.
  • Apply a single PEARL or OL25CX label near the cap with a unique descriptive ID.
  • Pack vials continuously (no gaps), mark the 4 corner IDs, and match the Excel Request Form 1:1.
  • Disclose expected high concentrations (e.g., biogas, soil profile N2O, or 13C-enriched CO2) beforehand.

✘ Common Mistakes to Avoid (Don't)

  • Don't stack multiple labels or use thick/permanent third-party stickers.
  • Don't submit 16 mm OD vials or factory pre-evacuated Labco vials.
  • Don't leave empty holes in trays or include unsubmitted sample IDs on the Excel form.
  • Don't wrap plumber's tape around vial caps.
  • Don't use ≥21G needles or pull syringes immediately without allowing pressure equilibration.
  • Don't drop off boxes labeled only "Samples" without your name, contact info, and emailed form.
Materials & Methods Reference

Analytical Instrumentation & Acknowledgments

Need instrument details for your thesis, manuscript, or conference poster? Use the specifications below and contact PEARL if you need the detailed GC spec sheet for your specific run date.

Greenhouse Gas & Biogas Chromatography

  • Standard GHG System: Bruker 450-GC (Varian/Bruker/Scion platform) equipped with Electron Capture Detector (ECD) for N2O, Flame Ionization Detector (FID) for CH4, and Thermal Conductivity Detector (TCD) for CO2 and O2.
  • Typical GC Detection Limits: ~25 ppm CO2, ~0.05 ppm CH4, ~0.03 ppm N2O (typical N2O analytical uncertainty ~0.01 ppm).
  • Calibration Standards: Certified multi-point reference gases supplied by Praxair / Linde; extended calibrations up to 100% N2O available on request.
  • Biogas / Rumen System: Agilent 490 Micro GC (measures H2, O2, N2, CH4, and CO2). Note: PEARL uses an Agilent 490 Micro GC, not an Agilent 6890.

Isotopic Spectroscopy & Partner Facilities

  • 15N2O Isotopomer CRDS: Picarro G5131-i Isotopic N2O Analyzer (Picarro Inc., Santa Clara, CA) for δ15N2O, δ15Nα, δ15Nβ, and δ18O.
  • δ13C-CO2 CRDS: Picarro Cavity Ring-Down Spectrometer for headspace δ13C-CO2 and CO2 concentration (>300 ppm CO2 required).
  • Partner IRMS Laboratory: Bulk 15N-N2O and 15N-N2 analyses are performed in collaboration with the Soil Science Stable Isotope Lab (Room 5E80, isoFlow GHG IRMS).
  • Thesis Examples: USask HARVEST repository theses detailing PEARL GC methodology include Example Thesis 1 and Example Thesis 2.

Preferred Manuscript & Poster Acknowledgment

Gas sample analyses were performed at the Prairie Environmental Agronomy Research Laboratory (PEARL), Department of Soil Science, University of Saskatchewan, Saskatoon, SK, Canada.

Common Questions

Frequently Asked Questions

Answers to the administrative, technical, and logistics questions most frequently asked by USask and external researchers.

What company/vendor name should external universities put on a Purchase Order (PO)?

Use "University of Saskatchewan" (not "BMO Soil Analytical Lab"). Contact Dr. Shannon Brown (shannon.brown@usask.ca) or lab5e19@gmail.com to obtain an official quotation for your PO. Invoices for external clients are typically issued quarterly for samples analyzed up to that point, or on custom dates if required by grant deadlines.

How do internal University of Saskatchewan researchers pay for analysis?

Provide a valid USask CFOAPAL account number on your Sample Submission / Request Form. Billing is processed internally based on the number of analyzed samples.

Does measuring CO2 and CH4 in addition to N2O cost extra?

No. Standard GHG analysis ($4.00/sample) simultaneously measures N2O, CO2, CH4, and O2 at no extra cost. Additional charges apply only for extremely high concentrations (or dedicated Biogas Micro GC runs at $8.00/sample), δ13C-CO2 CRDS runs ($14.00/sample), or batches of fewer than 100 samples.

Can I use my own Exetainer vials instead of PEARL-supplied vials?

Yes, provided they match our autosampler rack geometry: 12 mL glass vials measuring 15.6 mm OD × ~100–101 mm tall with cap (or Labco 5.9–6.8 mL vials if arranged in advance) with double septa and a single layer of paper or plastic label. Vials measuring 16 mm OD will not fit. Do not use factory pre-evacuated Labco vials—we recommend evacuating in-house or letting PEARL ship you freshly evacuated vials. We return external labs' own vials after analysis.

How long can I keep evacuated vials before sampling, and filled vials before shipping?

Before sampling: Fill evacuated vials within 2–4 weeks of receiving them. For multi-month studies, request staggered vial shipments rather than receiving all vials upfront.

After sampling: Filled double-septa vials generally remain stable for ~1 month (and often longer), but longer storage times increase the risk of occasional leaky-vial outliers. Aim to ship batches of 100–400 samples every 2–4 weeks.

What is the typical turnaround time, and are there seasonal closures?

During normal operation, PEARL aims to analyze samples within ~1 week of receipt and emails results immediately upon completion. However, turnaround times can extend to several weeks during peak field seasons, scheduled lab shutdowns, or University winter holiday closures (shipping/receiving typically closes in late December through early January). Always check availability in advance before shipping.

Does PEARL calculate gas fluxes or cumulative emissions?

Typically no—PEARL delivers raw/calibrated gas concentrations (ppm), and flux calculations are performed by the researcher as part of their study or thesis (flux calculation support involves an additional fee). We recommend the Global Research Alliance N2O chamber flux calculation guidelines by Venterea et al. (2020).

What happens if I don't use all the vials PEARL sent me?

Please return all PEARL-supplied vials (both filled and unused spares) after your project concludes, and remove any uncollected/empty sample IDs from your submission form. Missing or unreturned vials—as well as empty slots left listed on a submission form—are subject to billing.

Research Impact & Literature

Selected Scientific Publications Referencing PEARL Gas Analysis

Peer-reviewed journal articles utilizing gas chromatography, cavity ring-down spectroscopy, and greenhouse gas flux measurements conducted through PEARL / Department of Soil Science at the University of Saskatchewan, plus key methodological references recommended by the lab.

Showing 15 of 15 publications

  1. Livestock & Rumen CH4 2026

    Schreiner, B. A., Pereira, M. C. S., Lardner, H. A., Ribeiro, G. O., & Penner, G. B. (2026). The comparison of a barley-based complex forage mixture relative to a barley monoculture on yield and quality, and on dry matter intake, growth, and enteric methane emissions for beef heifers. Canadian Journal of Animal Science.

    USask forage and enteric methane gas analysis collaboration.

    DOI: 10.1139/cjas-2026-0036 ↗
  2. Livestock & Rumen CH4 2026

    Schreiner, B. A., Penner, G. B., Lardner, H. A., & Ribeiro, G. O. (2026). In vitro ruminal fermentation, nutrient degradability, methane production, and microbial protein synthesis of a barley-based complex forage mixture compared to a barley monoculture when harvested at two maturities. Canadian Journal of Animal Science.

    In vitro ruminal headspace CH4 & gas chromatography analysis at USask.

    DOI: 10.1139/cjas-2026-0037 ↗
  3. Isotopes & Pathways 2025

    Shorunke, A. T., Helgason, B. L., & Farrell, R. E. (2025). Evidence of the need for crop-specific N2O emission factors. Soil Biology and Biochemistry, 202, 109694.

    15N- and 13C-labelled crop residue incubation, GC concentration, and stable isotope tracing at PEARL, USask.

    DOI: 10.1016/j.soilbio.2024.109694 ↗
  4. Livestock & Rumen CH4 2025

    Liu, J., Knight, J. D., Lemke, R. L., Lardner, H. A., & Farrell, R. E. (2025). Diet and landscape controls on greenhouse gas emissions from cattle excreta in a semi-arid environment. Agriculture, Ecosystems & Environment, 382, 109469.

    Chamber N2O, CH4, and CO2 gas analysis from semi-arid prairie pastures at USask.

    DOI: 10.1016/j.agee.2025.109469 ↗
  5. Soil GHG & Cropping 2025

    Hangs, R. D., Schoenau, J. J., Knight, J. D., & Farrell, R. E. (2025). Variable rate precision application of feedlot cattle manure mitigates soil greenhouse gas emissions. Geoderma, 454, 117172.

    Field-scale soil GHG flux quantification via PEARL gas chromatography.

    DOI: 10.1016/j.geoderma.2025.117172 ↗
  6. Soil GHG & Cropping 2024

    Ferland, D., Wagner-Riddle, C., Brown, S. E., Bourgault, M., Helgason, W., Farrell, R. E., & Congreves, K. A. (2024). Improved nitrogen fertilizer management reduces nitrous oxide emissions in a northern Prairie cropland. Science of The Total Environment, 956, 177211.

    Joint USask & University of Guelph study evaluating N management and prairie N2O emissions.

    DOI: 10.1016/j.scitotenv.2024.177211 ↗
  7. Soil GHG & Cropping 2024

    Wagner-Riddle, C., Congreves, K. A., Brown, S. E., Helgason, W. D., & Farrell, R. E. (2024). Overwinter and spring thaw nitrous oxide fluxes in a northern Prairie cropland are limited but a significant proportion of annual emissions. Global Biogeochemical Cycles, 38, e2023GB008051.

    Year-round prairie N2O flux dynamics led by USask and University of Guelph collaborators.

    DOI: 10.1029/2023GB008051 ↗
  8. Soil GHG & Cropping 2024

    Pelster, D. E., Matteau, J.-P., Farrell, R., & Hernandez Ramirez, G. (2024). Tillage effects on growing season nitrous oxide emissions in Canadian cropland soils. Canadian Journal of Soil Science, 103(3), 344–352.

    Multi-site synthesis incorporating Canadian prairie N2O chamber datasets.

    DOI: 10.1139/cjss-2023-0075 ↗
  9. Livestock & Rumen CH4 2023

    Lemke, R. L., Machado, P. V. F., Alemu, A. W., Baron, V. S., Farrell, R. E., Iwassa, A. D., Liang, B. C., & Schoenau, J. J. (2023). Pastures under dry climate can be cold spots for soil nitrous oxide emissions from grazing animal excreta. Agriculture, Ecosystems & Environment, 357, 108684.

    Multi-site grazing excreta N2O chamber study analyzed in collaboration with USask.

    DOI: 10.1016/j.agee.2023.108684 ↗
  10. Isotopes & Pathways 2021

    Machado, P. V. F., Farrell, R. E., Deen, W., Voroney, R. P., Congreves, K. A., & Wagner-Riddle, C. (2021). Contribution of crop residue, soil, and fertilizer nitrogen to nitrous oxide emissions varies with long-term crop rotation and tillage. Science of The Total Environment, 767, 145107.

    15N-N2O source partitioning collaboration between University of Guelph and USask.

    DOI: 10.1016/j.scitotenv.2021.145107 ↗
  11. Soil GHG & Cropping 2021

    Liu, L., Knight, J. D., Lemke, R. L., & Farrell, R. E. (2021). Type of pulse crop included in a 2-year rotation with wheat affects total N2O loss and intensity. Biology and Fertility of Soils, 57, 699–713.

    Static-chamber N2O concentration analysis in Saskatchewan pulse–wheat rotations.

    DOI: 10.1007/s00374-021-01562-4 ↗
  12. Isotopes & Pathways 2019

    Congreves, K. A., Phan, T., & Farrell, R. E. (2019). A new look at an old concept: using 15N2O isotopomers to understand the relationship between soil moisture and N2O production pathways. SOIL, 5(2), 265–274.

    PEARL / agGAL Bruker 450-GC paired with Picarro G5131-i isotopic N2O analyzer in Exetainer vials.

    DOI: 10.5194/soil-5-265-2019 ↗
  13. Soil GHG & Cropping 2018

    David, C., Lemke, R., Helgason, W., & Farrell, R. E. (2018). Current inventory approach overestimates the effect of irrigated crop management on soil-derived greenhouse gas emissions in the semi-arid Canadian Prairies. Agricultural Water Management, 208, 19–32.

    High-frequency chamber N2O, CO2, and CH4 gas chromatography in Saskatchewan irrigated croplands.

    DOI: 10.1016/j.agwat.2018.06.006 ↗
  14. Wetlands & Landscapes 2010

    Pennock, D. J., Yates, T. T., Bedard-Haughn, A., Phipps, K., Farrell, R. E., & McDougal, R. (2010). Landscape controls on N2O and CH4 emissions from freshwater mineral soil wetlands of the Canadian Prairie Pothole Region. Geoderma, 155(3–4), 308–319.

    Foundational USask Soil Science chamber N2O and CH4 analysis across Prairie Pothole wetlands.

    DOI: 10.1016/j.geoderma.2009.12.015 ↗
  15. Methodology Guidelines 2020

    Venterea, R. T., Petersen, S. O., de Klein, C. A. M., Pedersen, A. R., Noble, A. D. L., Rees, R. M., Gamble, J. D., & Parkin, T. B. (2020). Global Research Alliance N2O chamber methodology guidelines: Flux calculations. Journal of Environmental Quality, 49(5), 1141–1155.

    Recommended by PEARL for researchers calculating chamber and microcosm fluxes from concentration data.

    DOI: 10.1002/jeq2.20118 ↗
Get Started

Contact, Shipping & Billing

Reach out before your experiment begins so we can schedule your evacuated vial deliveries, verify sample compatibility, and plan analytical capacity.

Lab Team & Contacts

  • Gas Analysis & Sample Submissions Frank Krijnen, Research Technician
    lab5e19@gmail.com
  • PEARL Lab Director & Quotes Dr. Shannon E. Brown, Assistant Professor
    shannon.brown@usask.ca • (306) 966-2703
  • Vials & Field Supplies Support Mark Cooke
    mwc044@mail.usask.ca
  • Founder / Professor Emeritus Dr. Richard E. Farrell, Professor Emeritus
    Department of Soil Science, USask
  • Partner IRMS Facility (Room 5E80) Kim Janzen, Lab Manager (kim.janzen@usask.ca)

Shipping & Campus Drop-Off

University of Saskatchewan
Attn: Lab 5E19 (PEARL)
51 Campus Drive
Saskatoon, SK S7N 5A8
Canada
  • USask On-Campus Drop-Off Room 5E19 (or leave outside Room 5E15 if nobody is in 5E15, 5E19, or 5E31).
  • USask Vial Pickup Arranged vials and labels are placed outside Room 5E31 with your name attached.

Billing & Purchase Orders

  • Internal USask Clients Include your valid CFOAPAL account number directly on the GHG Analysis Request Form.
  • External Institutions Request a quote before shipping. Issue Purchase Orders to vendor name: "University of Saskatchewan".
  • Invoicing Schedule Invoices are issued quarterly for analyzed samples, or earlier upon request to meet grant fiscal deadlines.
  • Official Department Pricing USask Soil Science Facilities & Rates ↗